İstenen sayfanın yerel versiyonuna yönlendirildiniz

AB-096

Determination of mercury at the rotating gold electrode by anodic stripping voltammetry


Summary

This Application Bulletin describes the determination of mercury by anodic stripping voltammetry (ASV) at the rotating gold electrode. With a deposition time of 90 s, the calibration curve is linear from 0.4 to 15 μg/L; the limit of quantification is 0.4 μg/L.

The method has primarily been drawn up for investigating water samples. After appropriate digestion, the determination of mercury is possible even in samples with a high load of organic substances (wastewater, food and semi-luxuries, biological fluids, pharmaceuticals).

İletişim

Metrohm Turkey

Ayazağa Mah. Azerbaycan Cad. No. 3I
34396 İstanbul

İletişim

Uygulamaları Bulun
Aramanızı geliştirin(641)

Bu uygulama şurada etiketlenmiştir

// Food – dairy// Water – seawater// Water – process water// Water – drinking water, tap water// Food – plant sources// Water – power plant water// Food – animal sources// Beverages – nonalcoholic// Pharmaceutical solutions// Water – wastewater// Water – others// Food – liquid food, functional food