You have been redirected to your local version of the requested page

AB-096

Determination of mercury at the rotating gold electrode by anodic stripping voltammetry


Summary

This Application Bulletin describes the determination of mercury by anodic stripping voltammetry (ASV) at the rotating gold electrode. With a deposition time of 90 s, the calibration curve is linear from 0.4 to 15 μg/L; the limit of quantification is 0.4 μg/L.

The method has primarily been drawn up for investigating water samples. After appropriate digestion, the determination of mercury is possible even in samples with a high load of organic substances (wastewater, food and semi-luxuries, biological fluids, pharmaceuticals).

Contact

Metrohm Romania

Str. Emil Racoviță nr. 25
041753 București

Contact

Find Applications
Filtrați-vă căutarea(641)

Această aplicație a fost clasificată sub

// Food – dairy// Water – seawater// Water – process water// Water – drinking water, tap water// Food – plant sources// Water – power plant water// Food – animal sources// Beverages – nonalcoholic// Pharmaceutical solutions// Water – wastewater// Water – others// Food – liquid food, functional food