Determination of mercury at the rotating gold electrode by anodic stripping voltammetry


概要

This Application Bulletin describes the determination of mercury by anodic stripping voltammetry (ASV) at the rotating gold electrode. With a deposition time of 90 s, the calibration curve is linear from 0.4 to 15 μg/L; the limit of quantification is 0.4 μg/L.

The method has primarily been drawn up for investigating water samples. After appropriate digestion, the determination of mercury is possible even in samples with a high load of organic substances (wastewater, food and semi-luxuries, biological fluids, pharmaceuticals).

お問い合わせ

メトロームジャパン株式会社

143-0006 東京都大田区平和島6-1-1
東京流通センター アネックス9階

お問い合わせ

アプリケーション検索
もう一度再検索(631)

このアプリケーションはこちらのタグが付いています

// Food – dairy// Water – seawater// Water – process water// Water – drinking water, tap water// Food – plant sources// Water – power plant water// Food – animal sources// Beverages – nonalcoholic// Pharmaceutical solutions// Water – wastewater// Water – others// Food – liquid food, functional food