요청하신 페이지의 적용되는 국가로 리디렉션되었습니다

AB-422

Determination of mercury in water using scTRACE Gold


Summary

This Application Bulletin describes the determination of inorganic mercury in water samples by anodic stripping voltammetry using the scTRACE Gold sensor. With a deposition time of 90 s, calibration is linear up to a concentration of 30 µg/L; the limit of detection lies at 0.5 µg/L.

문의하기

Metrohm AG

Ionenstrasse
9100 Herisau

문의하기