요청하신 페이지의 적용되는 국가로 리디렉션되었습니다

AB-342

Automated determination of calcium and magnesium in milk with thermometric titration


Summary

This Application Bulletin discusses the automated determination of calcium and magnesium in commercially available finished milk products. Calcium and magnesium in milk can be rapidly and easily titrated thermometrically using a standard solution of Na4EDTA as titrant.

The milk sample is first treated with a trichloroacetic acid solution to coagulate protein and fat and liberate calcium and magnesium as dissociated ions. The serum containing the calcium and magnesium is separated by filtration or centrifugation. An aliquot of milk serum is treated with acetylacetone to enhance the separation of the calcium and magnesium endpoints. The solution is then buffered with NH3/NH4Cl solution before titration. Thermometric titrations are conducted under conditions of constant titrant addition rate. Results are reported as mg Ca and Mg/100 mL.

문의하기

Metrohm AG

Ionenstrasse
9100 Herisau

문의하기

애플리케이션 찾기
선택 후 재검색

이 애플리케이션의 태그는 다음과 같습니다

// Food – dairy