요청하신 페이지의 적용되는 국가로 리디렉션되었습니다

AB-437

Installation instructions TitrIC flex II


Summary

The TitrIC flex II system is used for the fully automatic analysis of water samples using direct measurement, titration, and ion chromatography (IC). The following parameters are determined within a very short time: temperature, conductivity, pH, acid capacity, and in parallel, the concentrations of individual anions and cations with the resulting water hardness and ion balance. Other Metrohm instruments can be incorporated in the existing system at any time and used to measure additional parameters.


Analytical sequence

The sample is transferred from the first Pick&Place Module to two 930 Compact IC Flex instruments for the anion and cation analysis. The conductivity is measured at the same time. Following this measurement, the exact sample volume is transferred into the sample beaker of the second Pick&Place. Afterwards the temperature, pH, and acid capacity (p and m values) of the sample are determined.

The entire procedure is controlled by the MagIC Net software. The user enters the sample position and sample identification only into MagIC Net. All liquid handling steps as well as the titration part are done by OMNIS. MagIC Net triggers the individual sequences in OMNIS and receives the titration results via RS-232.

All data is summarized in a joint database containing all results by MagIC Net as well as the numerical values from OMNIS.

문의하기

Metrohm Korea Ltd.

서울특별시 송파구 정의로 8길, 9, 5층 (AJ 본관)
05836

문의하기

애플리케이션 찾기
선택 후 재검색

이 애플리케이션의 태그는 다음과 같습니다

// Water – drinking water, tap water