요청하신 페이지의 적용되는 국가로 리디렉션되었습니다

AB-223

Fully automated determination of uranium


Summary

This Bulletin describes the fully automated determination of uranium according to the method of Davies and Gray: Uranium(VI) is reduced in concentrated phosphoric acid solution with iron(II) to form Uranium(IV). With molybdenum as a catalyst, the excess iron(II) is oxidized with nitric acid. The nitrous acid that is formed is destroyed with sulfamic acid before uranium(IV) is titrated with a potassium dichromate solution in the presence of a vanadium catalyst.

문의하기

Metrohm Korea Ltd.

서울특별시 송파구 정의로 8길, 9, 5층 (AJ 본관)
05836