AB-004

Biamperometric titration method for the determination of antimony in lead


Summary

An automatic titration method is described using biamperometric endpoint indication for the determination of antimony in antimony-alloyed cable-sleeving lead (approx. 1% Sb). A 0.01 mol/L KBrO3 solution is used as the titrant.