Raman vs SERS… What’s the Difference?
1 févr. 2021
Article
If you’ve ever had a conversation with a Raman spectroscopist about the feasibility of a low-concentration sensing application, chances are you’ve heard them say “well, Raman may not be sensitive enough…but maybe SERS will work!” But what’s the actual difference between these two techniques, and why is SERS (surface-enhanced Raman scattering, or alternatively surface-enhanced Raman spectroscopy) recommended for low-concentration applications? Let’s explore the technical differences between Raman and SERS spectroscopies, as well as some of the practical considerations for how we regard the data for each.
In normal Raman spectroscopy, a laser source is incident directly on a sample (Figure 1a). The laser light is scattered by the bonds of the analyte, and the inelastically scattered light is collected and processed into a Raman spectrum. The non-destructive nature of the technique, the selectivity of Raman bands, and the insensitivity to water make Raman a useful analytical tool for both qualitative and quantitative studies of both organic and inorganic systems.
However, for decades Raman spectroscopy was an underutilized technique in real-world applications. This can be attributed to its two major limitations: 1) the inherent insensitivity of Raman, as only ~1 in 106 incident photons are Raman scattered; and 2) fluorescence emission interference, which depends on the nature of the analyte molecule and the excitation wavelength used. Fluorescence is a competing phenomenon that is much more efficient than Raman scattering, and can thus completely overwhelm the Raman signal.
Though they depend on the scattering strength of the analyte molecule and the sample matrix in question, typical limits of detection for normal Raman scattering can range from ~1–10% in concentration. For certain applications such as disease detection or narcotics identification, this limit may be several orders of magnitude higher than what is required! In this case, an application scientist might recommend a SERS measurement. The hardware required would be the same as for a normal Raman measurement, but different sampling is required for SERS analysis. To understand the difference, let’s discuss a bit about the SERS effect.
In the 1970's, several research groups observed that the Raman signal from organic molecules like pyridine was greatly enhanced when adsorbed to a roughened metallic substrate (Figure 1b) [1–3]. While several theories emerged to account for this observation, it is today generally accepted that the mechanism for enhancement is two-fold: the electromagnetic enhancement mechanism accounts for the dominant contribution, while a chemical mechanism accounts for a smaller portion of the enhancement.