AN-NIR-019
Human stool analysis by near-infrared spectroscopy
Determination of moisture, fat, and nitrogen
Introduction
Several diseases like pancreatic insufficiency or hepatic disorders cause malabsorption or maldigestion. The resulting changes in stool composition (e.g., in moisture, nitrogen content, or fat content) provide important information for medical diagnostics.
Time-consuming lab methods which require a significant amount of sample preparation can be replaced by near-infrared spectroscopy (NIRS) for fast screening of human stool samples with no sample preparation. The NIRS method is easy to use – no chemicals are required and results are given for several parameters within one minute.
Experimental equipment
522 human stool samples were analyzed on a Metrohm DS2500 Solid Analyzer with a modified DS2500 Holder for petri dishes (Figure 1). Stool samples were positioned into the petri dishes for the analysis in diffuse reflection mode. Reference values for moisture, fat, and nitrogen content were obtained with the respective primary methods.
Equipment | Metrohm number |
---|---|
DS2500 Solid Analyzer | 2.922.0010 |
DS2500 Holder | 6.7430.040 |
Vision Air 2.0 Complete | 6.6072.208 |
Results
The obtained Vis-NIR spectra (Figure 2) were used to create prediction models for the different reference parameters. The data set was split into calibration and validation sets to verify the quality of the prediction models. Correlation diagrams which display the relation between the Vis-NIR prediction and the reference values are shown in Figures 3–5 together with the respective figures of merit (FOM).
Result moisture in human stool
Result fat in human stool
Result nitrogen in human stool
Conclusion
This Application Note shows the feasibility of NIR spectroscopy for the analysis of moisture, fat, and nitrogen content in human stool samples by NIRS, which can be conducted without chemicals or sample preparation.