AN-H-097

Determination of calcium and magnesium in harvested salt


Summary

A measured amount of salt is titrated directly with a solution of 1 mol/L tetrasodium EDTA to thermometrically determined endpoints for Ca and Mg. Acetylacetone is added to alter the Ca and Mg EDTA stability constants for better endpoint sharpness.