Chuyển hướng tin nhắn

AN-V-109

Selenium in wastewater after UV digestion


Tóm tắt

Selenium is determined by cathodic stripping voltammetry (CSV) at the hanging mercury drop electrode (HMDE). Se(IV) is deposited on the surface of the mercury drop in sulfuric acid electrolyte under addition of copper ions as Cu xSe y.

Wastewater samples containing organic contaminants have to be digested by UV irradiation before analysis. In addition, the sample has to undergo a second irradiation step at pH 7−9 to reduce Se(VI) to Se(IV), since only Se(IV) is electrochemically active.

 

 

 

Liên hệ

CÔNG TY TNHH METROHM VIỆT NAM

Tòa nhà Park IX, số 08 Đường Phan Đình Giót, Phường 2, Quận Tân Bình
Thành phố Hồ Chí Minh

Liên hệ