Chuyển hướng tin nhắn

AN-C-158

Column stability of the Metrosep C 6 - 250/4.0 with Inline Ultrafiltration and Inline Eluent Preparation


Tóm tắt

The column stability of the Metrosep C 6 - 250/4.0 was determined in long-term laboratory tests. Two injection series per day were run on each of six days in a row. Each series was comprised of nine tap water injections, three check standard injections and six tap water injections. The IC system was shut down on the seventh day of each series. As a whole, the system ran over 10 weeks and counted a total of 2,150 injections. The results show an outstanding reproducibility and verify the high column stability.

Liên hệ

CÔNG TY TNHH METROHM VIỆT NAM

Tòa nhà Park IX, số 08 Đường Phan Đình Giót, Phường 2, Quận Tân Bình
Thành phố Hồ Chí Minh

Liên hệ