Chuyển hướng tin nhắn

AB-241

Determination of cadmium and lead by anodic stripping voltammetry at a mercury film electrode


Tóm tắt

This Application Bulletin describes the determination of cadmium and lead at a mercury film electrode (MFE) by anodic stripping voltammetry (ASV). The mercury film is plated ex situ on a glassy carbon electrode and can be used for up to one day. With a deposition time of 30 s, the limit of detection is ß(Cd2+) = 0.02 µg/L and ß(Pb2+) = 0.05 µg/L. The linear working range for both elements goes up to approx. 50 μg/L using the same deposition time.

Liên hệ

CÔNG TY TNHH METROHM VIỆT NAM

Tòa nhà Park IX, số 08 Đường Phan Đình Giót, Phường 2, Quận Tân Bình
Thành phố Hồ Chí Minh

Liên hệ