İstenen sayfanın yerel versiyonuna yönlendirildiniz

AN-V-174

Arsenic in zinc plant electrolyte


Summary

The concentration of As(total) in zinc plant electrolyte is determined by anodic stripping voltammetry (ASV) on a lateral gold electrode in HCl electrolyte. Due to the high excess of zinc in the sample the deposition potential has to be adapted. A second potential approx. 100 mV more negative than the arsenic signal has to be applied to selectively oxidize interfering antimony. For sample preparation the sample was passed through a cation exchange column to reduce the concentration of zinc in the measuring solution.
İletişim

Metrohm Turkey

Ayazağa Mah. Azerbaycan Cad. No. 3I
34396 İstanbul

İletişim