You have been redirected to your local version of the requested page

8.000.6063

Post-column chemistry for improved optical absorption detection


Summary

UV/VIS detection is one of the most sensitive detection techniques in trace-level chromatography. Sometimes, however, spectrophotometric detection lacks sensitivity, selectivity or reproducibility and chemical derivatizations are required. By using Metrohm`s rugged and versatile flow-through reactor, single- or multi-step derivatizations can be done fully automatically, in either pre- or post-column mode at any temperature between 25…120 °C. The variable reactor geometry allows to adjust the reactor residence time of the reactants according to derivatization kinetics. The flexibility of the reactor is demonstrated by optimizing four widespread post-column techniques: the relatively slow ninhydrin reaction with amino acids and the fast derivatizations of silicate, bromate and chromate(VI).
Контакт

Metrohm RUS

ул. Садовники д.3А
115487 Москва

Контакт

Find Applications
Фильтровать поиск

Это применение было помечено как

// Textiles, leather// Water – process water// Amino acids, cellulose, lignin, proteins// Water – others