AN-V-238

2024-04

Bismuth stabilizer in an electroless Ni plating bath

Straightforward determination over a wide concentration range with the scTRACE Gold


Summary

The electroless nickel plating process is known for its superior surface finishing properties with excellent resistance to wear and corrosion. The efficiency of the ENIG (electroless nickel, immersion gold) and ENEPIG (electroless nickel, electroless palladium, immersion gold) processes in manufacturing printed circuit boards strongly depends on the exact composition of the electroless nickel plating bath. Monitoring stabilizer concentrations in Ni baths (e.g., bismuth, Bi) is crucial for high-quality coatings.

Anodic stripping voltammetry (ASV) allows fast and robust monitoring of the bismuth concentration in Ni plating bath samples. This determination is performed on the scTRACE Gold. It is a combined sensor containing working, reference, and auxiliary electrodes integrated on a single ceramic substrate. It does not need extensive maintenance such as mechanical polishing. Measurements can be performed with the 884 Professional VA. This method is suited for manual or automated systems.


Sample

Electroless Ni plating bath


Experimental

884 Professional VA, fully automated for voltammetric  analysis.
Figure 1. 884 Professional VA, fully automated for voltammetric analysis.

Add water, the electroless Ni plating bath sample, and the supporting electrolyte into the measuring vessel. The determination of bismuth is carried out with the 884 Professional VA (Figure 1) using the parameters specified in Table 1. The concentration is determined by two additions of a bismuth standard addition solution.

The scTRACE Gold is electrochemically activated prior to the first determination.

Table 1. Parameters
Parameter Setting
Mode DP – Differential Pulse
Deposition potential -0.1 V
Deposition time 30 s
Start potential 0.0 V
End potential 0.3 V
Peak potential Bi 0.15 V

Electrodes

  • scTRACE Gold

Results

Determination of bismuth in an electroless Ni bath  sample containing 1 mg/L Bi (30 s deposition time, sample volume  25 μL in 10 mL water).
Figure 2. Determination of bismuth in an electroless Ni bath sample containing 1 mg/L Bi (30 s deposition time, sample volume 25 μL in 10 mL water).

The typical Bi concentration in nickel plating bath samples is around 1 mg/L. However, samples containing 100 μg/L Bi can be reliably determined with a 30 s deposition time using this method. The scTRACE Gold sensor quickly measures bismuth in Ni baths over a broad concentration range.

Table 2. Result
Sample Bi (mg/L)
Electroless Ni plating bath containing 1 mg/L Bi 1.07