요청하신 페이지의 적용되는 국가로 리디렉션되었습니다

AN-C-117

MiPT – Metrohm intelligent Partial Loop Technique


Summary

Calibration of lithium, sodium, ammonium, zinc, potassium, magnesium, and calcium using the partial loop technique and cation chromatography with direct conductivity detection. This technique allows a calibration range of 1:100 (e.g. 1 μg/L to 100 μg/L corresponding to 2 μL to 200 μL injected volume) out of 1 calibration solution. Applying the full range of partial loop injection to the samples one calibration covers a sample concentration range of 1 to 10'000 e.g. 2 μL of a 10 mg/L solution corresponds to the highest calibration level (100 μg/L) while 200 μL of a 1 μg/L solution corresponds to the lowest calibration level.

문의하기

Metrohm Korea Ltd.

서울특별시 송파구 정의로 8길, 9, 5층 (AJ 본관)
05836

문의하기