Determination of ferrous ion in acidic solutions with ceric solution


Summary

This Application Note looks at the determination of ferrous ion in acidic solutions from approximately 0.25 g/L by thermometric titration with ceric titrant. The exothermic oxidation reaction shows a sharp endpoint that is detected using the Thermoprobe as a sensitive temperature sensor.