You have been redirected to your local version of the requested page

The addition of dyes to provide uniform coloration and enhance visual appeal in food products is a common practice. Rhodamine B is a dye utilized extensively in biotechnology and industrial applications and is one of several colorants banned for use as food additives in Europe and North America. The most common analytical methods for detection of illicit dyes in food products, GC/MS and HPLC, are laboratory-based instrumental methods that require specialized training. With Misa (Metrohm Instant SERS Analyzer), detection of trace amounts of Rhodamine B in ground cayenne pepper is quick and easy after a facile extraction procedure with minimal material consumption. Rhodamine B can be detected in cayenne powder at a concentration of 50 μg/g. However, a simple concentration step improves that limit to 10 μg/g.

Ground cayenne pepper bought commercially was doped with Rhodamine B (RhB) and tested with Misa to simulate a realistic food screening scenario.

Gold NP SERS standard reference spectrum of Rhodamine B.
Figure 1. Gold NP SERS standard reference spectrum of Rhodamine B.

To establish a reference spectrum, a pure RhB standard (50 μg/g in ultrapure water) was analyzed using gold nanoparticles (Au NPs). The unique SERS spectrum shown in Figure 1 can be used to create a library entry for RhB.

Addition of sample solution to the vial for measurement

A stock solution of RhB in methanol was prepared. Purchased cayenne powder was treated with serial dilutions of the stock (3 mL stock to 1 g cayenne) to yield samples with 1000, 500, 100, 50, 10, 5, and 1 μg/g of RhB. Samples were thoroughly mixed and air-dried. To prepare extracts, 0.1 g of each spiked sample was added to a vial with 400 μL of methanol, shaken to mix, and left to settle for 10 minutes. To prepare test samples, 50 μL of the methanol extract was pipetted into a vial with 400 μL of Au NP solution and 50 μL of 0.5 mol/L salt solution. The vial was shaken to mix, and then placed into the vial attachment on Misa for testing.

Table 1. Experimental Parameters
Instrument Acquisition
Firmware 0.9.33 Laser Power 5
Software Misa Cal V1.0.15 Int. Time 10 s
Misa Vial Attachment 6.07505.040 Averages 10
ID Kit - Au NP 6.07506.440 Raster ON

In Figure 2, overlaid spectra of RhB indicate detection down to 50 μg/g. For each concentration tested, the baseline spectrum from unadulterated cayenne was subtracted from the average of baseline-corrected, replicate measurements.

Figure 2. Gold NP SERS concentration profile of RhB extracted from adulterated cayenne powder. Spectra are baselined, with Au NP and control subtracted.

To improve trace detection and spectral signal-to-noise, a very simple concentration method was applied to each extract. All extracts were fully air-dried, then resuspended in methanol to yield a 5x increase in concentration. The spectra in Figure 3 demonstrate detection of RhB down to 10 μg/g.

Figure 3. RhB profile after 5x concentration.

Detection of Rhodamine B in the field

Using the large end of the scoop, add 3–4 scoops of sample to a 2 mL vial. Add methanol to the vial until ~1/3 full. Cap and shake the vial gently to mix, then let sample rest for 2 minutes. Fill a clean vial halfway full with Au NPs. Using pipettes, add 1 drop each of sample solution and NaCl solution to the Au NPs, then cap and shake the vial gently to mix. Insert into vial attachment on Misa for measurement.

Table 2. Requirements for Field Test Protocol
ID Kit - Au NP 6.07506.440
includes: Gold nanoparticles (Au NP)
Scoop
Disposable pipettes
2 mL glass vials
Reagents  
Methanol  
NaCl solution 3 g NaCl in 100 mL water
Test settings Use ID Kit OP on MISA

Trace levels of detection, ease of sample preparation, and rapid assay times collectively recommend Misa as a reliable, cost-effective solution for high-throughput, on-site identification of adulterated food products.

Contact

Metrohm Belgium nv

Drabstraat 198
2550 Kontich

Contact