Du er blevet omdirigeret til din lokale version af den anmodede side

AN-H-119

Determination of ferric ion in acidic solutions


Summary

This Application Note deals with the determination of ferric ion in acidic and copper-free solutions using thermometric titration. The ferric ion is reduced by iodide. The released iodine reacts exothermically when titrated with thiosulfate solution. The endpoint is determined through temperature plotting by the temperature sensor Thermoprobe.