AN-C-167

Quality test of an automatic and direct 1:2000 dilution applying Inline Dilution Techique


Summary

Automatic dilution reduces manual work and improves the reproducibility and accuracy of the results. So far, the Inline Dilution Technique (MIDT) has been tested in a range of max. 1:100. Using a dedicated sample needle this range is enlarged significantly. This AN shows the performance of an Inline Dilution with a factor of 1:2000 as well as a comparison of manual and inline dilution for a dilution factor of 1:1000.