Du er blevet omdirigeret til din lokale version af den anmodede side

AB-131

Determination of aluminum by adsorptive stripping voltammetry


Summary

This Application Bulletin describes a voltammetric method for the determination of aluminum in water samples, dialysis solutions, sodium chloride solutions and digestion solutions (e.g. of lyophilisates). The method utilizes the complexation of the Al3+ ion by Calcon (Eriochrome blue black R). The formed complex can easily be reduced electrochemically at 60 °C. The limit of quantitation depends on the purity of the reagents used and is approx. 5 µg/L.

Kontakt

Metrohm Nordic ApS

Lyskær 3C. 1. th.
2730 Herlev

Kontakt