Aplicações
- AB-130Chloride titrations with potentiometric indication
Besides acid-base titrations, the titrimetric determination of chloride is one of the most frequently used titrimetric methods of analysis. It is employed more or less frequently in practically every laboratory. This Bulletin shows you how to determine chloride in a wide range of concentrations using automatic titrators. Silver nitrate is normally used as titrant (for environmental reasons one should refrain from using mercury nitrate). The titrant concentration depends on the chloride content of the sample to be analyzed. It is crucial to choose the correct electrode for samples with low chloride contents.
- AB-070Polarographic determination of nitrate in water samples, soil and plant extracts, vegetable juices, meat and sausages, fertilizers, liquid manure, etc.
The photometric determination of nitrate is limited by the fact that the respective methods (salicylic acid, brucine, 2,6-dimethyl phenol, Nesslers reagent after reduction of nitrate to ammonium) are subject to interferences. The direct potentiometric determination using an ion-selective nitrate electrode causes problems in the presence of fairly large amounts of chloride or organic compounds with carboxyl groups. The polarographic method, on the other hand, is not only more rapid, but also practically insensitive to chemical interference, thus ensuring more accurate results. The limit of quantification depends on the matrix of the sample and is approximately 1 mg/L.
- AB-137Coulometric water content determination according to Karl Fischer
This Application Bulletin gives an overview of the coulometric water content determination according to Karl Fischer. Amongst others, it describes the handling of electrodes, samples, and water standards. The described procedures and parameters comply with the ASTM E1064.
- AB-143Determination of complexing agents in detergents
Application Bulletin AB-076 contains a description of the polarographic determination of low concentrations (1–100 mg/L) of NTA and EDTA in bodies of water. NTA, EDTA and citrate have gained in importance as complexing agents and builders due to the fact that the laws of some countries have made it necessary to find a substitute for phosphates in detergents. This Bulletin describes the determination of larger quantities of complexing agents in detergents using potentiometric titration. The ion-selective copper electrode (Cu-ISE) is used here as the indicator electrode. The determination of complexing agents is not disturbed by the other constituents often present in detergents.
- AB-196Determination of formaldehyde by polarography
Formaldehyde can be determined reductively at the DME. Depending on the sample composition it may be possible to determine the formaldehyde directly in the sample. If interferences occur then sample preparation may be necessary, e.g. absorption, extraction, or distillation.Two methods are described. In the first method formaldehyde is reduced directly in alkaline solution. Higher concentrations of alkaline or alkaline earth metals interfere. In such cases the second method can be applied. Formaldehyde is derivatized with hydrazine forming the hydrazone, which can be measured polarographically in acidic solution.
- AB-316Determination of phosphoric acid in liquid fertilizer with 859 Titrotherm
The phosphoric acid content can be easily titrated with a standardized solution of 2 mol/L NaOH. The interfering calcium content in phosphoric fertilizer can be eliminated by adding a saturated oxalate solution.
- AB-027Potentiometric titration of chloride and bromide in the presence of each other
If chloride and bromide are present in approximately equal molar concentrations they can be titrated directly with silver nitrate solution after addition of barium acetate. If, however, the molar ratio n(Br-) : n(Cl-) changes from 1 : 1 to 1 : 5, 1 : 10, 5 : 1 or 10 : 1 then greater relative errors must be expected with this method. The Bulletin describes an additional titration method that allows bromide to be determined in the presence of a large excess of chloride. The determination of small chloride concentrations in the presence of a large excess of bromide is not possible by titration.
- AB-209Water in insulating oils, hydrocarbons, and their products – Accurate and reliable determination by Karl Fischer titration
Only coulometric Karl Fischer titration can determine low water contents with sufficient accuracy. This Application Bulletin describes the direct determination according to ASTM D6304, ASTM E1064, ASTM D1533, ASTM D3401, ASTM D4928, EN IEC 60814, EN ISO 12937, ISO 10337, DIN 51777, and GB/T 11146. The oven technique is described according to ASTM D6304, EN IEC 60814, and DIN 51777.
- AB-230Potentiometric determination of nonionic surfactants based on polyoxyethylene adducts with the NIO electrode
The titrimetric determination of nonionic surfactants on the basis of polyoxyethylene adducts (POE adducts) is described in the Bulletin. The basis for the determination is the transfer of the nonionic surfactant into a pseudo-cation compound and its precipitation titration with sodium tetraphenylborate (Na-TPB). The NIO electrode is used for the indication of the potentiometric titration. This Bulletin describes determinations in raw products, formulations and wastewater and draws attention to special features, possibilities, limits and disruptions.
- AB-407Automated volumetric Karl Fischer titration
This Application Bulletin provides information regarding the MATi 10 (Metrohm Automated Titration) system. MATi 10 is a completely configured system for automatic volumetric Karl Fischer titration with which the water content in liquid and solid samples can be determined. Up to 24 samples can be analyzed directly in 75 mL titration vessels. The samples are weighed into the titration vessels and covered with an aluminum foil. This prevents falsification of the water content.